Nuclear Medicine AtlasInteractive atlas

Acquisition Parameters: SPECT & Planar

Representative adult activity, collimator, energy window, matrix, and timing by study

Evidence B#reference#procedure#SPECT#planar#quantitationUpdated 2026-07-27 · Physician-reviewed
High-yield mode — key points, high-yield facts, tables & quizzes only. Toggle off for full text.

Quick Answer

This is a representative adult acquisition reference for common SPECT/planar studies — radiopharmaceutical, typical activity, collimator, energy window, matrix/acquisition, and timing. Values are representative and institution/guideline-dependent (SNMMI/EANM procedure standards); pediatric activity is weight-based (EANM Dosage Card / North American consensus). The three levers that most often decide image quality are the collimator (matched to isotope energy), the energy window (centered on the photopeak), and adequate counts/time.

⚠️ Representative parameters for orientation and study — verify against current society guidelines and your local protocol before clinical use. Activities in mCi (MBq in parentheses where helpful).

Collimator Selection (the First Decision)

Isotope (photopeak) Collimator
Tc-99m (140 keV), Tl-201 (~70 keV), I-123 (159 keV) Low-energy (LEHR / LEAP)
In-111 (173/247), Ga-67 (93/185/300) Medium-energy
I-131 (364 keV) High-energy

A low-energy collimator with a high-energy isotope → septal-penetration (star) artifact. Pinhole collimation is used for thyroid/parathyroid and small-organ high-resolution imaging.

Renal, Genitourinary

Study Agent Activity Collimator / window Acquisition Timing / notes
Dynamic renography Tc-99m-MAG3 ~5–10 mCi LEHR / 140 keV Dynamic 64×64: flow 1–3 s frames, then 15–20 s ×20–30 min; supine, posterior Diuretic F+20/F−15/F0; hydrate, empty bladder
GFR Tc-99m-DTPA ~3–5 mCi LEHR / 140 Gates camera method or plasma sampling Filtered marker
Cortical / DMSA Tc-99m-DMSA ~2–5 mCi LEHR or pinhole / 140 Static high-count (~300–500k) planar (post + obliques); or pinhole Image ~2–4 h; scarring at ~6 mo
Cystography (direct) Tc-99m-pertechnetate/SC per protocol (intravesical) LEHR / 140 Continuous dynamic through fill/void Low gonadal dose

Skeletal

Study Agent Activity Collimator / window Acquisition Timing
Bone scan (WB) Tc-99m-MDP/HDP ~20–30 mCi (740–1110 MBq) LEHR / 140 WB planar 256×1024 (~10–15 cm/min) ± SPECT 128×128, 60–120 views, 20–40 s/view Delayed 2–4 h; hydrate/void
Three-phase Tc-99m-MDP as above LEHR / 140 Flow (1–3 s), blood pool (~5 min), delayed Infection/CRPS/viability

Cardiac

Study Agent Activity Collimator / window Acquisition Notes
MPI (Tc-99m) sestamibi/tetrofosmin stress/rest ~10–30 mCi (1- or 2-day) LEHR / 140 Gated SPECT 64×64, 180° (RAO→LPO), ~60–64 views, 20–25 s/view, 8–16 gates Image 15–60 min post; prone/upright, AC
MPI (Tl-201) thallous chloride ~3–4 mCi LEHR / 70 keV (+167) Stress–redistribution ± reinjection Redistribution = viability
MUGA Tc-99m-RBC ~20–25 mCi LEHR / 140 Gated planar LAO best-septal, 16–24 frames/cycle, high counts In-vitro labeling; stable rhythm
PYP (ATTR) Tc-99m-PYP ~15–20 mCi LEHR / 140 Planar + SPECT at 1 h (± 3 h) Monoclonal screen; SPECT mandatory

Pulmonary

Study Agent Activity Collimator / window Acquisition Notes
Perfusion (Q) Tc-99m-MAA ~3–5 mCi (200k–500k particles) LEHR / 140 8 planar views ± SPECT; supine injection Reduce particles if R-to-L shunt
Ventilation Technegas / DTPA aerosol / Xe-133 per agent LEHR Wash-in/out (Xe) or single-breath/equilibrium Central deposition in COPD (aerosol)

Hepatobiliary & GI

Study Agent Activity Collimator / window Acquisition Notes
HIDA Tc-99m-mebrofenin ~3–8 mCi (↑ if high bilirubin) LEHR / 140 Dynamic 60 min (1 min/frame), anterior; delayed 3–4 h/24 h Fast 4 h (not >24); morphine/CCK; phenobarbital (neonate)
Gastric emptying Tc-99m-SC (egg meal) ~0.5–1 mCi LEHR / 140 Anterior+posterior at 0,1,2,4 h; geometric mean Standard meal; image to 4 h
GI bleed Tc-99m-RBC ~20–25 mCi LEHR / 140 Dynamic 60–90 min (1 min/frame) + delayed In-vitro label; appears-and-moves
Meckel Tc-99m-pertechnetate ~5–10 mCi (peds weight-based) LEHR / 140 Dynamic + statics 30–60 min, anterior Cimetidine pretreat

Endocrine

Study Agent Activity Collimator / window Acquisition Notes
Thyroid uptake/scan I-123 (or Tc-99m-pertechnetate) I-123 ~100–400 µCi; Tc ~2–10 mCi Pinhole / 159 (I-123) or 140 (Tc) Planar pinhole; uptake at 4–6 & 24 h (I-123) Trapping (Tc) vs organification (I)
Parathyroid Tc-99m-sestamibi ~20–25 mCi LEHR/pinhole / 140 Dual-phase (10–15 min & 1.5–2.5 h) + SPECT/CT ± ¹⁸F-fluorocholine PET
MIBG (I-123) I-123-MIBG ~10 mCi LEHR / 159 Planar + SPECT at 24 h Thyroid block; hold interfering drugs
SSTR (legacy) In-111-pentetreotide ~6 mCi Medium-energy / 173+247 Planar + SPECT at 4 & 24 h (±48) Superseded by DOTATATE PET

Infection / Lymphatic / CNS

Study Agent Activity Collimator / window Acquisition Notes
Labeled WBC Tc-99m-HMPAO WBC ~7–10 mCi LEHR / 140 Planar ± SPECT/CT at 1–4 h Pair with sulfur-colloid marrow
" In-111-oxine WBC ~0.5 mCi Medium-energy / 173+247 Imaging at 24 h Cleaner abdomen; delayed imaging
Sentinel node Tc-99m-SC (filtered)/tilmanocept ~0.5–2 mCi LEHR / 140 Dynamic + static ± SPECT/CT Filter colloid; intraop probe
DaTscan I-123-ioflupane ~3–5 mCi LEHR/fanbeam / 159 SPECT at 3–6 h Thyroid block; review meds
Cisternography In-111-DTPA (intrathecal) ~0.5 mCi Medium-energy / 173+247 Sequential 1,3,6,24,48,72 h NPH pattern; pledgets for leak
Brain perfusion Tc-99m-HMPAO/ECD ~20 mCi LEHR / 140 SPECT; inject at seizure onset (ictal) Fixed-at-injection snapshot

Board-Level Synthesis

Three levers decide a SPECT/planar study: the collimator (energy-matched — LEHR for Tc-99m/Tl-201/I-123, medium-energy for In-111/Ga-67, high-energy for I-131), the energy window (centered on the photopeak), and adequate counts/time. A low-energy collimator with I-131 produces a septal-penetration star artifact.

Timing is part of the protocol: bone delayed 2–4 h, HIDA 60 min (delayed to 4/24 h), gastric emptying to 4 h, DMSA scarring re-imaged at ~6 months, In-111 studies at 24 h (± 48). Cardiac and PYP require SPECT (gating for MPI; myocardial confirmation for PYP).

Activities are representative and guideline/institution-dependent; pediatric activity is weight-based with a minimum diagnostic floor (EANM Dosage Card / North American consensus). Special techniques recur: pinhole for thyroid/parathyroid, filtered colloid for sentinel node, in-vitro RBC labeling for bleed/MUGA, and diuretic/hydration/bladder control for renography.

  • Reference: Acquisition parameters — PET, therapy administration protocols, quantitative thresholds & decision cutoffs.
  • Physics: Radiation detection & instrumentation (collimators), quality control procedures.

Figure / Diagram Suggestions

  • A collimator-by-isotope selection chart with the star-artifact warning.
  • A one-page protocol card grid (agent / activity / collimator / window / timing).

Self-Check

Q1. Which collimator is required for I-131 imaging, and what happens if a low-energy one is used?

Answer: A high-energy collimator (I-131 photopeak 364 keV); a low-energy collimator causes septal-penetration (star) artifacts.

Q2. Why is In-111 imaged with a medium-energy collimator and at ~24 hours?

Answer: In-111's photopeaks are 173/247 keV (medium-energy), and its 2.8-day half-life supports delayed (24 h) imaging (e.g. WBC, OctreoScan, cisternography).

Q3. What acquisition and timing features are mandatory for a PYP cardiac amyloid study?

Answer: SPECT (to confirm myocardial vs blood-pool/rib uptake) at 1 h (± 3 h), alongside a monoclonal-protein screen.

Q4. How is pediatric administered activity determined?

Answer: Weight-based with a minimum diagnostic activity, per the EANM Dosage Card / North American consensus — not adult activity scaled arbitrarily.

Evidence & sources

BSNMMI/EANM procedure standards — representative adult activities, collimator/energy-window selection, and acquisition/timing for planar and SPECT studies (verify locally).
BEANM Dosage Card / North American consensus — weight-based pediatric administered activities.
Cite this page. Nuclear Medicine Atlas. “Acquisition Parameters: SPECT & Planar.” v1.76, 2026-10-25. Permalink: https://nuc.med/acquisition-parameters-spect/ Report an issue
Nuclear Medicine Atlas · v1.76 · built 2026-10-25Terms · Privacy · Disclaimer