Nuclear Medicine AtlasInteractive atlas

Quality Control Procedures

Gamma camera, SPECT, PET/CT, dose calibrator, and generator QC, cadence and pass criteria

Evidence B#instrumentation#physics#quality-control#basic-principlesUpdated 2026-07-27 · Physician-reviewed
High-yield mode — key points, high-yield facts, tables & quizzes only. Toggle off for full text.

Quick Answer

A defined quality-control program keeps imaging quantitative and detects failures before they mimic disease. The essentials: gamma camera — daily uniformity and energy peak, periodic spatial resolution/linearity (bar phantom) and SPECT center-of-rotation (COR); PET/CT — daily normalization/blank and cross-calibration; dose calibrator — daily constancy, periodic linearity, accuracy, geometry; generator — Mo-99 breakthrough (≤ 0.15 µCi/mCi) and aluminum (≤ 10 µg/mL). Each test has a cadence and an acceptance criterion; failing one has a characteristic artifact.

Gamma Camera (SPECT) QC

Test Cadence What it detects / criterion
Energy peak (photopeak) Daily Correct window centering; drift causes count loss/artifact
Uniformity (flood) Daily Non-uniform detector response → ring/patchy SPECT artifacts; integral/differential uniformity (commonly ≤ ~5% for CFOV)
Spatial resolution & linearity (bar phantom) Weekly Four-quadrant bar phantom; smallest resolved bars = resolution; bar straightness = linearity
Center of rotation (COR) Weekly–monthly SPECT reconstruction alignment; COR error → "tuning-fork"/doughnut ring artifacts and blurring
Tomographic uniformity / sensitivity Periodic System sensitivity (cpm/µCi) and SPECT uniformity

Extrinsic floods include the collimator (test the clinical system); intrinsic floods are without the collimator (test the detector). A non-uniformity projects as a ring artifact at the corresponding radius on SPECT — the single most important reason daily uniformity is mandatory.

PET/CT QC

Test Cadence Purpose
Daily normalization / blank scan Daily Detector-block stability; drift degrades uniformity/quantitation
Calibration / well-counter cross-calibration Periodic Keeps SUV quantitative (scanner ↔ dose calibrator ↔ clock)
CT QC Per CT program Attenuation-map accuracy, CT number/water
NEMA NU-2 performance Acceptance/annual Resolution, sensitivity, scatter fraction, NECR, image quality

A miscalibrated cross-calibration (scanner, dose calibrator, and clock must agree) silently biases SUV — a quantitation error, not a visual one.

Dose Calibrator QC

The instrument behind every administered activity — its QC is non-negotiable:

Test Cadence Method / criterion
Constancy Daily Long-lived reference source (Cs-137, Co-57); reading stable within ~±5–10%
Accuracy Annual (and at install) NIST-traceable source(s); within ~±5% (±10% action)
Linearity Quarterly Decay method or shield (Calicheck) method across the activity range
Geometry At installation Volume/container-dependence correction factors

Well Counter & Survey Instruments

  • Well counter: daily constancy and energy calibration; efficiency for quantitative counting (blood dosimetry, wipe tests).
  • Survey / GM meters: annual calibration, daily battery/constancy check; used for contamination and exposure surveys and package/area monitoring.

Generator & Radiopharmaceutical QC

Test Limit Note
Mo-99 breakthrough ≤ 0.15 µCi Mo-99 / mCi Tc-99m Radionuclidic purity, at administration
Aluminum (Al³⁺) ≤ 10 µg/mL Column breakthrough (colorimetric strip); excess disrupts labeling/colloids
Radiochemical purity typically ≥ 90–95% Chromatography; free-pertechnetate / hydrolyzed-reduced species
Sterility/endotoxin Per USP Compounded and PET products

Failure → Artifact Map

QC failure Characteristic artifact
Non-uniformity Ring artifact on SPECT (at the defect radius)
COR error Doughnut / tuning-fork blurring, point-source ring
Wrong photopeak/energy Count loss, degraded contrast
Collimator–isotope mismatch Septal-penetration star (e.g. low-energy collimator with I-131)
Cross-calibration error Biased SUV (silent quantitative error)
Mo-99 breakthrough Added dose, degraded images, high-energy contamination

Board-Level Synthesis

The daily anchors are gamma-camera uniformity + energy peak, PET normalization/blank, and dose-calibrator constancy. The periodic anchors are bar-phantom resolution/linearity and SPECT COR (weekly–monthly), and dose-calibrator linearity (quarterly)/accuracy (annual)/geometry (install).

A non-uniformity projects as a ring artifact at the corresponding SPECT radius, and a COR error causes doughnut/tuning-fork blurring — the two failures most likely to mimic or obscure disease, and the reason their QC is frequent.

Two limits anchor generator radionuclidic/chemical purity: Mo-99 breakthrough ≤ 0.15 µCi/mCi and aluminum ≤ 10 µg/mL. The dose calibrator underlies every administered activity — daily constancy, quarterly linearity, annual accuracy, and geometry at install — and PET cross-calibration (scanner ↔ dose calibrator ↔ clock) keeps SUV trustworthy, a quantitation error that is invisible on the image.

  • Physics: Radiation detection & instrumentation, counting statistics & ROC (chi-square/constancy), radiopharmaceuticals & production (generator QC).
  • Pitfalls: Pearls, pitfalls & normal variants (artifacts).

Figure / Diagram Suggestions

  • A QC cadence calendar (daily/weekly/quarterly/annual by instrument).
  • A failure → artifact plate (ring, doughnut, septal-penetration star).
  • A dose-calibrator four-test schematic (constancy/accuracy/linearity/geometry).

Self-Check

Q1. Which gamma-camera QC tests are performed daily, and what artifact does a uniformity failure cause?

Answer: Daily uniformity (flood) and energy-peak checks; a non-uniformity projects as a ring artifact at the corresponding radius on SPECT.

Q2. What are the four dose-calibrator QC tests and their cadences?

Answer: Constancy (daily), linearity (quarterly), accuracy (annual), and geometry (at installation).

Q3. State the Mo-99 breakthrough and aluminum limits for a Tc-99m generator eluate.

Answer: Mo-99 ≤ 0.15 µCi per mCi Tc-99m (at administration) and aluminum ≤ 10 µg/mL.

Q4. A SPECT study shows a doughnut/tuning-fork blurring artifact. Which QC parameter is likely off?

Answer: The center of rotation (COR) — misalignment causes point-source ring/doughnut and tomographic blurring.

Evidence & sources

BSNMMI / IAEA / ACR QC programs — gamma-camera uniformity/COR/bar-phantom, PET normalization, dose-calibrator constancy/linearity/accuracy/geometry.
BUS NRC / USP — Mo-99 breakthrough (0.15) and aluminum (10 µg/mL) limits; radiochemical-purity testing.
Cite this page. Nuclear Medicine Atlas. “Quality Control Procedures.” v1.76, 2026-10-25. Permalink: https://nuc.med/qc-procedures/ Report an issue